System Specifications

Summary of performance targets and operating envelope for the Automated Laser Beam Characterizer instrument. See full details in Documents.

Input Beam Specifications

ParameterValue
WavelengthAt least 2 µm
Max Power1 - 10 – 10 W
Minimum Power1 - 50 – 50 mW
Beam Diameter4 - 35 – 35 mm

Measurement Specifications

MeasurementAccuracy
Normalized Stokes Parameters±0.05 for all parameters
Wavelength±10 nm
M² factor±0.1 up to M²=3
Beam Divergence±10 µrad
Total Powerless than 5 %

Component Specifications

ComponentParameterSpecificationUnits
QWPPhase Delay~π/2 @ 2 µmwaves
QWPClear Aperture> 42mm
Linear PolarizerExtinction Ratio< -20dB
Linear PolarizerClear Aperture> 42mm
Power MeterAperture Size4 – 10mm
Power MeterPower Intake*1 – 1000mW
ND FilterOD Level2N/A
LensSpectral RangeAt least 2µm
LensEFFL600 – 1000mm
LensClear Aperture> 42mm
Beam SplitterSplit Ratio50/50N/A
Beam SplitterPolarization effectsLowN/A

* Power level of Power Meter can be less than system input power when ND filter is applied.

Notes

  • Measurement algorithms follow ISO 11146 guidance where applicable.
  • Divergence and waist derived from z-scan; M² via far-field slope or quadratic fit.
  • Polarization via rotating QWP + fixed LP; Stokes vector computed and normalized.